Feature(may vary by option.)

●【Product Introduction】: FFU HEPA H14 filter laminar flow hood has CE and ISO certification. The HEPA H14 filtered laminar flow hood ensures that work surfaces are constantly filled with air in a laminar flow manner, and the updated fan filter unit captures 99.99% of particles larger and smaller than 0.3 microns in diameter, with filtration efficiency meeting 100/IS0 Level 5 cleanliness.
●【Product Material】: FFU HEPA H14 filter laminar flow hood is made of galvanized frame, aluminum fan blades, pure copper motor and thickened metal casing, using a safety lock to ensure that the fan filter unit works more durable and safer and efficient.
●【Product Parameters】: 22.6”x22.6”x10.23”, the filtered air flow speed can be set to three speeds, the filtered air flow is 0.45m/s (±20%), and the air volume is 600- 700m3/H (depending on the working environment)
●【Scope of Application】: The fan filter laminar flow hood supports 110V American plug and can be used in dust-free clean rooms, laboratories, tissue culture seedlings, laminating machines (mobile phone film), laboratories for mobile phone LCD display repair, etc.
●【Product After-Sale】:We will continue to improve the details of our products to make them more perfect! Each FFUHEPA H14 filter laminar flow hood has been carefully inspected and packaged before leaving the factory. Thank you for your use, thank you!

[(accessory) filter]




[HEPA-Filter Laminar Flow Hood]





Description

laminar flow hoodlaminar flow hood
Updated Fan Filter UnitUpdated Fan Filter Unit
Laminar Air Flow HoodLaminar Air Flow Hood

Product parameters:

Size: 22.6*22.6*10.23 inches

Cabinet material: galvanized plate

Wheel blade material: aluminum blade

Filtration efficiency: 100/ISO level 5

Wind speed: 0.45m/s±20%

Air volume: 600-700m3/h (depending on the working environment)

Cleanliness: Class 100 (US Federal Standard 209E)

Noise: Low speed 41-51. Medium speed 52-59 High speed 59-65

Vibration value: ≤0.45m/s2

Voltage: AC110V, 50HZ, single phase

Power: 1200W

Speed ​​regulation mode: three gear speed regulation

Wiring method: American standard three plug

Maintenance:

1. Do not open and close the filter repeatedly.

2. Do not operate the machine with wet hands, otherwise there is a danger of electric shock.

3. Do not block the intake valve with any other objects. The fan filter laminar flow hood needs to be kept away from heaters or other heat sources.

4. Wipe the laminar air supply unit frequently to keep it clean.

5. The fan filter laminar flow hood should be replaced regularly (per cycle every 8-12 months): After replacing the high-efficiency filter, the laminar air supply should be re-tested The technical parameters of the unit can only continue to work after meeting the requirements.

Detection:

It is recommended to use a particle counter regularly to detect dust particles in the working area of ​​the fan filter laminar flow hood in accordance with the US federal standard The number (cleanliness) and whether the filter is leaking.