Product parameters:
Size: 22.6*22.6*10.23 inches
Cabinet material: galvanized plate
Wheel blade material: aluminum blade
Filtration efficiency: 100/ISO level 5
Wind speed: 0.45m/s±20%
Air volume: 600-700m3/h (depending on the working environment)
Cleanliness: Class 100 (US Federal Standard 209E)
Noise: Low speed 41-51. Medium speed 52-59 High speed 59-65
Vibration value: ≤0.45m/s2
Voltage: AC110V, 50HZ, single phase
Power: 1200W
Speed regulation mode: three gear speed regulation
Wiring method: American standard three plug
Maintenance:
1. Do not open and close the filter repeatedly.
2. Do not operate the machine with wet hands, otherwise there is a danger of electric shock.
3. Do not block the intake valve with any other objects. The fan filter laminar flow hood needs to be kept away from heaters or other heat sources.
4. Wipe the laminar air supply unit frequently to keep it clean.
5. The fan filter laminar flow hood should be replaced regularly (per cycle every 8-12 months): After replacing the high-efficiency filter, the laminar air supply should be re-tested The technical parameters of the unit can only continue to work after meeting the requirements.
Detection:
It is recommended to use a particle counter regularly to detect dust particles in the working area of the fan filter laminar flow hood in accordance with the US federal standard The number (cleanliness) and whether the filter is leaking.