Calibration Made Easy How to calibrate a pH meter?
1. Before calibration, prepare 4 beakers with 250ml of clean water/pH 4.00/pH 6.86/pH 9.18 standard solution.
2. Rinse the probe electrode with clean distilled water pH0 and wipe dry. Turn on the pH meter and adjust to pH mode.
3. Immerse the electrode in pH 6.86 buffer solution and when the value stabilizes, press the "CAL" key for 5 seconds, the screen will display "CAL" and when pH 6.86 blinks 3 times, the calibration of pH 6.86 is complete.
4. Repeat above steps to calibrate pH4.00 or pH9.18